EDITORIAL
Year : 2008 | Volume
: 19 | Issue : 6 | Page : 895--902
Biofeedback systems and adaptive control hemodialysis treatment
Ahmad Taher Azar Biomedical Engineering Department, Misr University for Science & Technology, 6th October City, Egypt
Correspondence Address:
Ahmad Taher Azar Ahmad Orabi Square, Azar Building, Menouf, Menoufeya Egypt
On-line monitoring devices to control functions such as volume, body temperature, and ultrafiltration, were considered more toys than real tools for routine clinical application. However, bio-feedback blood volume controlled hemodialysis (HD) is now possible in routine dialysis, allowing the delivery of a more physiologically acceptable treatment. This system has proved to reduce the incidence of intra-HD hypotension episodes significantly. Ionic dialysance and the patient«SQ»s plasma conductivity can be calculated easily from on-line measurements at two different steps of dialysate conductivity. A bio-feedback system has been devised to calculate the patient«SQ»s plasma conductivity and modulate the conductivity of the dialysate continuously in order to achieve a desired end-dialysis patient plasma conductivity corresponding to a desired enddialysis plasma sodium concentration. Another bio-feedback system can control the body temperature by measuring it at the arterial and venous lines of the extra-corporeal circuit, and then modulating the dialysate temperature in order to stabilize the patients«SQ» temperature at constant values that result in improved intra-HD cardiovascular stability. The module can also be used to quantify vascular access recirculation. Finally, the simultaneous computer control of ultrafiltration has proven the most effective means for automatic blood pressure stabilization during hemodialysis treatment. The application of fuzzy logic in the blood-pressure-guided biofeedback control of ultrafiltration during hemodialysis is able to minimize HD-induced hypotension. In conclusion, online monitoring and adaptive control of the patient during the dialysis session using the bio-feedback systems is expected to render the process of renal replacement therapy more physiological and less eventful.
How to cite this article:
Azar AT. Biofeedback systems and adaptive control hemodialysis treatment.Saudi J Kidney Dis Transpl 2008;19:895-902
|
How to cite this URL:
Azar AT. Biofeedback systems and adaptive control hemodialysis treatment. Saudi J Kidney Dis Transpl [serial online] 2008 [cited 2021 Jan 24 ];19:895-902
Available from: https://www.sjkdt.org/article.asp?issn=1319-2442;year=2008;volume=19;issue=6;spage=895;epage=902;aulast=Azar;type=0 |
|
|